Response Time Measurement

 

Response Time Measurement Unit Of Temperature Sensor

 

The equipment measures the response time of the thermocouple or platinum resistance sensors, which is one of the critical parameters, in the case of temperature sensors. The range of the unit is 999.999 Seconds with 1 millisecond as resolution. A crystal controlled clock is used for measurement of time for better accuracy, stability and reliability. The continuous display of input is provided for monitoring purposes.

The unit is capable of accepting

• 0 – 40.00 mille Volts for Thermocouple inputs.
• 0 – 400.00 ohms for platinum resistance sensors.

The input - either thermocouple of RTD is selected by switch. The four wire method is used for measurement of RTD input; however three wire or two wire method can also be implemented.

The instrument measures the time between the two set points preset with the help of two sets of keyboard and alphanumeric display, and is stored in Battery Backed up RAM RTC.

The measurement can be performed either on UP-GOING or DOWN-COMING modes, selectable by data entry. Time count is started, when the input of the sensor under test, goes above the minimum setting and is stopped, when the input goes above the maximum setting in UP-GOING mode. Similarly time count starts when input comes below upper settings, and stops when it comes below lower settings in case of DOWN-COMING mode.

The special features such as error indications (with audible output) on wrong settings, invalid inputs and reserve polarity connection in case of thermocouple input, make the instrument safe and reliable.

The special Diagnostic software is provided to check the proper functioning of equipment. One can enter into diagnostic mode with the help of DIG switch provided at rear panel.

The indication is given on 20x4 Alphanumeric backlit display for time and input.

The equipment works on 230 +\- 10% Volts, 50 Hz, single phase AC mains and has size of 350 mm (W) x 140 mm (H) x 270 mm (D)

 
 
 

Hot Junction Welder | Resistivity Meter | Response Time Measurement Unit | Capacitor Discharge fine wire Welding Machine | Capacitor Discharge Welding Machine | Time Synchronising System | Quality Testing Devices | Digital Safety Ohms Meter | Siren Controller System | Stress Relieving / Calibration / Type Testing – Cycling of